Single-Use Pump Sizing: Choosing the Right PIXER Model
Single-use pump sizing is usually treated as a catalogue exercise: find the flow rate you need, pick the model whose headline number is just above it, and move on. With a fixed-displacement diaphragm pump it can be done properly in a few lines of arithmetic, and the same arithmetic tells you whether the model you picked will run in a sensible speed range.
This guide works through that calculation for the PIXER range from P0 to P40, then covers the other inputs that decide the model: viscosity, pressure, hold-up volume and the unit operation the pump feeds. Every figure comes from the MonoBiologics pump page and from Alphinity, the manufacturer of the PIXER platform.
Single-Use Pump Sizing Starts With Displacement
PIXER is a positive displacement diaphragm pump. Each revolution of the drive moves a fixed volume of fluid, so flow follows motor speed directly. The manufacturer describes the flow as backpressure-independent and a direct function of motor speed, which is what makes the sums below reliable.
The displacement per revolution is the number to start from. It rises from 0.73 mL on the P0 to 185 mL on the P40. Multiply it by the speed in revolutions per minute and you have the flow in millilitres per minute.
The platform covers flows from 0.73 mL/min to 240 L/min. That span is why the practical question is never whether a PIXER can reach your flow, only which of the five models reaches it at a comfortable speed.
PIXER Models at a Glance
The list below puts the five core models side by side. Two of them, the P0 and the P20, are also offered in Fast Prime versions.
- P0: 3 diaphragms, 0.73 mL per revolution, maximum flow 1.1 L/min at 1,500 RPM, internal volume 22 mL, tubing ID 1/8 in.
- P10: 5 diaphragms, 1.7 mL per revolution, maximum flow 5.1 L/min at 3,000 RPM, internal volume 45 mL, tubing ID 1/8 in. or 1/4 in.
- P20: 5 diaphragms, 16.7 mL per revolution, maximum flow 40.1 L/min at 2,400 RPM, internal volume 300 mL, tubing ID 1/2 in. or 3/4 in.
- P30: 5 diaphragms, 135 mL per revolution, maximum flow 175.5 L/min at 1,300 RPM, internal volume 1,300 mL, tubing ID 3/4 in. or 1 in.
- P40: 7 diaphragms, 185 mL per revolution, maximum flow 240.5 L/min at 1,300 RPM, internal volume 2,150 mL, tubing ID 1 in. or 1.5 in.
The range is built on odd diaphragm counts of three, five and seven. Because no two diaphragms oppose each other, the manufacturer reports smoother output than four-diaphragm designs, and independent testing is cited for 67% more flow at equivalent speed and the same target flow at 50% lower motor speed.
Working Out the Model From Your Flow Rate
The simplest rule is to choose the smallest model that covers your maximum flow. The upper limits to compare against are 1.1 L/min for the P0, 5.1 L/min for the P10, 40 L/min for the P20, 175 L/min for the P30 and 240 L/min for the P40.
Then check the speed. Divide the flow you need by the displacement per revolution of the candidate model.
- 3 L/min: 3,000 mL/min divided by 1.7 mL gives about 1,760 RPM on a P10, well inside its 1 to 3,000 RPM range.
- 20 L/min: 20,000 mL/min divided by 16.7 mL gives about 1,200 RPM on a P20, comfortably below its 2,400 RPM limit.
- 100 L/min: 100,000 mL/min divided by 135 mL gives about 740 RPM on a P30, which leaves a wide margin up to 1,300 RPM.
Compare the result with the nominal speed quoted for the model, which is 1,750 RPM for the P10 and P20 and 600 RPM for the P30. Running at or below nominal leaves headroom up to the maximum speed, which is useful if throughput rises later.
For a pump that will serve several unit operations, size to the highest flow it will see. Because the displacement is fixed, turning the speed down for a lower-flow step needs no change to the assembly.
Check Viscosity, Pressure and Hold-Up Volume
Flow is only half of the duty. The remaining checks are quick, and on the PIXER most of them are the same across the range.
- Viscosity: the pump handles fluids up to 300 cP, and flow is independent of viscosity. For context, the manufacturer notes that high-concentration monoclonal antibody formulations routinely exceed 50 cP and can reach 200 cP or more.
- Pressure: the operating pressure is up to 6 bar on every model.
- Temperature: the process fluid range is 4 to 40 °C.
- Sterilisation: the pump is compatible with gamma irradiation. It is not for steam or autoclave sterilisation.
- Materials: the body is polypropylene, with Novelast TPV diaphragms. Wetted materials are FDA compliant, USP Class VI and BSE/TSE free, and the pumps are made in ISO Class 7 cleanrooms.
Two behaviours of the pump matter for the layout. It is one-way by design, so either motor direction pumps forward and the pump acts as a built-in check valve; draining is handled by valves in the flow path. It also primes itself from a flooded top inlet, and has been tested lifting fluid from 3 metres below the inlet, although a top-fed set-up is still recommended.
Hold-up volume and product recovery
The internal volume of the pump head is the product held inside it, so it matters most when the batch is small or the product is valuable. It runs from 22 mL in the P0 to 2,150 mL in the P40, with 45 mL inside a P10.
The vertical, gravity-fed inlet floods the pump and lets it drain fully, which is the design feature behind the low hold-up. On a batch of a few litres, the difference between 22 mL and 300 mL of hold-up is worth weighing, which is one more reason to start from the smallest model that covers the flow.
Matching the Model to the Application
Once the numbers agree on a model, confirm it fits the unit operation. The PIXER pump page lists the main applications, and the sizing steps above apply to each.
- TFF and filtration feed: near-pulseless delivery keeps transmembrane pressure stable. For the wider pump choice on this step, see how diaphragm and peristaltic feed pumps compare for TFF.
- Cell and gene therapy: AAV, lentiviral and cell products suit a pump where the product never contacts moving parts.
- mRNA and LNP processing: the flow is low-shear.
- Inline dilution and dosing: the P10 and P20 take up to 10 injection ports, and the manufacturer lists mixing ratios from 1 to 99%.
- Scale-up and continuous manufacturing: behaviour is consistent from the P0 to the P40, so scale-up keeps the same pump technology from bench to manufacturing.
At the larger end, the P20 for mid-scale GMP work sits between development and full-scale manufacturing, while the P30 for manufacturing-scale buffer and media delivery handles large-volume transfer. Where the pump feeds a complete filtration skid, the single-use TFF system shows how it fits into a fully assembled set-up.
Sizing a PIXER Pump With MonoBiologics
MonoBiologics supplies the PIXER single-use pump range. The pump head is single-use and the drive unit is reusable, so there is no pump head to clean or revalidate between batches.
To size a pump, send us your flow rate, operating pressure and fluid. We will reply with the datasheet for the right model and, if you ask for it, a quotation. Custom outlet and injection port layouts are available.
Frequently Asked Questions
How do I size a single-use diaphragm pump?
Choose the smallest model whose maximum flow covers your requirement, then divide your flow by the model's displacement per revolution to find the motor speed. If that speed sits well inside the model's range, the choice is sound. Then confirm viscosity, pressure and temperature fall inside the limits for that pump.
Can a PIXER pump handle viscous fluids?
Yes, up to 300 cP. Because it is a positive displacement pump, flow does not depend on viscosity, so a viscous buffer or a concentrated antibody formulation is pumped at the same flow for a given motor speed. The manufacturer reports no output loss or cavitation across that range.
Is a PIXER pump suitable for shear-sensitive products?
It is designed for them. There is no tubing compression and no impeller, so the product never contacts moving parts. That avoids the pressure spikes and shear associated with tubing squeezed by rollers, which is why it is used for cell and gene therapy products, viral vectors and mRNA.
Can a PIXER pump run in reverse?
No. It is a one-way pump, and either motor direction pumps forward, so it acts as a built-in check valve. If a process needs to drain a line, that is handled by valves in the flow path rather than by reversing the pump.
Pump sizing comes down to displacement, speed and the limits of the fluid. Do the arithmetic first, choose the smallest model with comfortable headroom, and check that the viscosity, pressure and sterilisation route fit.
If you would like a second opinion on a flow requirement, send our team your flow, pressure and fluid and we will confirm the right PIXER model.